SPIE Ðóññêèé Óêðà¿íñüêà English
 
Fluorescent Diagnostics System of Oncological Diseases

Appointment

The system of fluorescence diagnosis of cancer (SFDOZ) designed for minimally invasive diagnosis of dysplasia in the clinical environment changes and malignancies of internal organs (gastrointestinal tract, female reproductive organs) in the early stages of using domestic photosensitizer Giperflav. Fluorescence diagnosis allows live differentiate healthy and diseased areas of the mucous internal organs for the timely appointment of conservative treatment to prevent their degeneration into cancer.

Specifications:
  ÑÔÄ.Ì6-1Ñ.2 ÑÔÄÃ
Emitter type diod laser GaInN
Spectral range, nm 460-490 440-470
Number of channels the emission 6 1
The radiation power at the distal end, MWt 2-3 8-10
The spectral range of the spectrometer, 350 nm 350 - 800 350 - 800
The range of registration of fluorescence, nm 500 - 800 500 - 800
The length of the fiber-optic carrier systems, m 5 3
The diameter of the distal tool, mm 2.8 3.0
Dimensions, mm:
block the radiation source 225 õ 135 õ 65 100 õ 50 õ 50
fiber-optic spectrometer 180 õ 135 õ 55 250 õ 140 õ 60
Weight, not more kg:
block the radiation source 0.5 0.5
fiber-optic spectrometer 1.0 1.5
Supply voltage V
block the radiation source 9 220
fiber-optic spectrometer 5 5

Developer: Laboratory of Optical Instrumentation, Medical Department of the optical and opto-electronic devices instrument-making faculty of NTU "KPI"
Supervisor: Ph.D. Denisov N.V. (Office tel.: 38-044-454-9482)